CQ6031/CQ6031C/CQ6051/CQ6051C

Features

.30MHz/50MHz, 2 channels,4 traces
.1mV/div~20V/div
.CRT readout
.cursor measurement, 10sets memory for front
panel setting save & recall
.Buzzer alarm & LED indicators
.ALT, MAG function(X5,X10,X20)
.CH1 signal output,Z-Axis input
.Hold-off function
.TV synchronization, X-Y mode
.Build-in 6 digital frequency counter (CQ6031C/
CQ605C)

 


CQ6051
Large view
 
             Technical Data   CQ6031/ CQ6031C/ CQ6051/ CQ6051C

CRT

Type 6-inch rectangular with internal graticule, 0%, 10%, 90%, and 100% marks
Z-axis input Coupling:DC
Sensitivity: 5V or more
Maximum input voltage:30V(DC+AC peak) at 1kHz or less
band width: DC~2MHz
input impedance: 33kΩ (50MHZ);47kΩ (30MHz)Ω
Accelerating voltage Apporx.2KV
Illumination Fron t panel control

VERTICAL SYSTEM

Sensitivity

1mV~20V/div, 14 step in 1-2-5 sequence
Vernier Vertical Sensitivity Continuously variable to 1/2.5 or less of panel-indicate value
Sensitivity Accuracy ≤3%;5mV~20V/div; ≤5%;1mV~2mV
bandwidth(-3B) DC~7Hz(1mV ~ 2mV/div);DC~30Hz/50MHz(5mV ~ 20V/div)
Rise Time     50ns(DC~7Hz);11.67ns(DC~30Hz);7ns(DC~50Hz)
Input Impedance 1MΩ±2% //approx.25pF
Max Input Voltage 400V (DC + AC peak) at 1kHz or less
Input Coupling AC, DC, GND
Vertical Mode CH1, CH2, DUAL(CHOP/ALT), ADD, CH2 INV

HORIZONTAL SYSTEM

Sweep Time
0.2us~0.5s / div;20 steps in 1-2-5 sequence
Sweep Time Accuracy ±3%(±5% at x5/x10MAG) ( ±8% at x 20MAG))
Sweep Magnificaton x5, x10, x20MAG
Max sweep time 20nS/DIV(10nS/DIV are uncalibrated)
ATL-MAG function Available

TRIGGER

Trigger Mode AUTO, NORM, TV
Trigger Source VERT-MODE,CH1, CH2, LINE, EXT
Trigger Coupling AC, HFR, LFR, TV-V(-), TV-H(-)
Trigger Slope "+" or "-" polarity
Trigger Sensitivity Mode Frequency INT EXT
AUTO and NORM 20Hz~5MHz(50MHz) 0.5 div 200mV
20Hz~2MHz(30MHz)
5Hz~40MHz(50MHz) 1.5 div 800mV
2Hz~20MHz(30MHz)
40Hz~50MHz(50MHz) 2.0 div 1 div
20Hz~30MHz(30MHz)
TV sync signal 1 div 200mVpp
Hold-off time

Input impedance:1MΩ±3%, 25pF±5pF

Max.External Input Voltage Max. input voltage: 400V (DC+ACpeak) at 1KHz
External Input Impedance 1MΩ±5% // approx.25pF

X-Y OPERATION

Sensitivity Accuracy 1mV~20V/div
X-axis Bandwidth DC~500KHz(-3dB)
Phase Error 3° or less from  DC~50kHz
Calibrator Output 1kHz Square wave, 2Vpp±2%
Bulid-in frequency counter    10Hz~30/50MHz; 6 digits

Power source                         AC 110/230V±10%, 50Hz/ 60Hz

Size                                        310(W) x 150(H) x 455(D)mm

WEIGHT                                 9kg
ACCESSORIES                       One operation manual, one fuse, one power cable, two probes