TEST INSTRUMENT CATEGORY
|
|
|
|
|
| |
|
CQ5010X
STUDENT DUAL CHANNEL OSCILLOSCOPE
Features
.10MHz
dual channel (Y and y)
.Y channel magnification (X5)
.TV synchronizing, X-Y operating
.Easy to operate
.High performance, sensitive5mV/DIV .Electrical
encoder switch to make high stability |

CQ5010X
|
| |
| Technical Data |
CQ5010X |
| CRT |
| Type |
3" round |
| Display area |
8 x 10DIV (1DIV=6mm) |
| Potential |
1.3kV |
| Lightering color |
Green |
| VERTICAL SYSTEM |
| Sensitivity |
5mV / DIV~5V/DIV±3%, Yˣ5MAG |
| Width of band (-3dB) |
DC: 0~10MHz AC:
10Hz~10MHz |
| Max. Input Voltage |
400V (DC + ACpeak) |
| Input Coupling |
AC, DC |
| Input Impedance |
1MΩ±3% 30pF±5pF |
| Trimming ratio |
2.5:1 |
| HORIZONTAL SYSTEM |
Sweep time
|
0.1s/ DIV~0.1μ s/ DIV±3% |
| Trimming ratio |
2.5:1 |
| TRIGGER SYSTEM |
| Mode |
AUTO, NORM, TV |
| Source |
INT(Y,y,ALT), LINE, EXT |
| Polarity |
"+"or"-" |
| Trigger sensitivity |
INT: 1DIV, EXT: 0.3V, TV: 2DIV |
| External trigger
input |
Input impedance: 1MΩ ±3% 25pF
±5pF
Max. input voltage: 160V(DC+ACpeak) |
| X-Y OPERATION |
| Sensitivity |
X: 0.5V/ DIV Y: 5mV/ DIV~5V/
DIV |
| Width of band (-3dB) |
DC: 0~1MHz AC: 10HHz-1MHz |
| Phase difference |
≤3° (DC~50kHz) |
| CALIBRATION |
| Source |
1kHz±2% 0.5Vp-p±2% square wave |
| POWER SOURCE |
110~127 VAC±10%, 220~240VAC±10% 50Hz±2Hz, 60Hz±2Hz |
ACCESSORIES
|
Operation manual x 1
Power cable x 1 ,Fuse x1
Probe x 2 |
| DIMENSIONS & WEIGHT |
140(W) x
196(H) x 290(D)mm
Approx 3kg |
|
|
|
|
Copyright © Shanghai MCP Corp, 2009. All Rights Reserved |
|
Add: Fl.3(East),Blk.4, Bai Da Industrial Park,No.1030 Tong Pu Rd., Shanghai, China